Customer Release Notes

WinCal 3.2 Specifics for Rohde & Schwarz VNA
============================================

WinCal 3.2 Rohde & Schwarz Driver

Support for the ZVK and ZVM analyzers from Rohde & Schwarz was included in WinCal 3.2. They are the same 
internally in respect to firmware and base hardware, but differ in top frequency.  They both are similar 
to most other VNAs and are four sampler models that are capable of using LRRM, SOLR and TRL calibrations.

The ZVK and ZVM are however different from most other VNAs in some aspects and special concerns are listed here.

Some specifics for the Rohde & Schwarz ZVK and ZVM:
---------------------------------------------------
* Minimum required firmware version is 3.52

* This instrument requires a "fake cal" to accept an external set of error correction terms.  
  That is why you will see a fast "cal" go by after WinCal has read the raw data it needs. 
  This "fake cal" is done at fastest possible sweep speed (max IF) and if the IF setting later 
  is lower than 300 Hz you will see "CA?" as an indication on the VNA screen of a cal being ON,
  as opposed to the normal "CAL".  This is not a problem. The cal is still valid.  To always 
  get the "CAL" indication you can turn off making the fake cal at max IF, but then you will 
  have to wait for WinCal's readings and the fake cal readings at the same IF setting (typically
  low sweep speed for stability)
  
* You can turn off fast "fake cal" by adding the line below in the Configuration section in drvna.ini  
    [Configuration] 
    RohdeSchwarzZVFastDummyCal=0
    
* The ZVK & ZVK have no calkit term for load inductance, so that value is not sent with a WinCal 
  probe calkit.  This only affects front panel calibrations.  WinCal SOLT is using the value.
  
* Dont use averaging during calibration.  The fake cal can easily time out.  It is better to 
  only work with IF bandwidth during calibration.